Salient Characteristics
LAB-17-004
Equipment must be able to automatically create anaerobic, microaerophilic, and capnophilic conditions in anaerobic jars.
Equipment must be small enough to fit on a table top.
Equipment must have the capability of using precise pressure measurements to control the evacuation and refilling of the anaerobic jar
Equipment must have capability of producing exact, repeatable conditions with gas mixtures that will always be within 0.5% of the required value.
Equipment must be able to detect any sign of leakage.
The vendor must provide an instrument training program to all staff that will use the instrument.
Please note we are looking for these salient characteristic or a Brand Name or Equal , Advanced Anoxomatî III