REQUEST FOR INFORMATION
1.0 DESCRIPTION
The Veterans Health Administration (VHA) Non-Expendable Equipment National Program identified Automated External Defibrillator (AED) as a candidate for VHA-wide (otherwise referred to as national ) single award Requirements contract. The Department of Veterans Affairs (DVA) Office of Acquisition Operations (OAO) Strategic Acquisition Center (SAC) on behalf of VHA is issuing this Request for Information (RFI) in accordance with the FAR 10.002(b)(2)to determine the availability and technical capabilities of qualified sources.
The SAC in conjunction with VHA is seeking a qualified source to manufacture ZOLLî brand name or equal AED on an agency wide basis.VA intends to award a Requirements Contract with Firm-Fixed Price (FFP) orders IAW FAR 16.503, Requirements Contract to a supplier of this equipment. Vendors will be required to deliver AED to VHA facilities throughout the United States, U.S. Territories, Guam, and the Philippines. Delivery requirements shall be specified under each individual order.
The proposed acquisition will consists of a 12 month base period and two (2) 12 month option periods from the date of award. However, the Government's decision as to whether or not to continue with the contract, upon annual review, will be based upon successful performance during each previous year and continued need.
The applicable North American Industry Classification System (NAICS) code is 334510, Electromedical and Electrotherapeutic Apparatus Manufacturing. All interested companies must be registered in the System for Award Management (SAM) prior to receiving an award. You may access SAM at https://www.sam.gov/portal/public/SAM/ . All Service Disabled Veteran Owned Small Businesses and Veteran Owned Small Business must be certified with the appropriate NAICS code on Vendor Information Pages at https://www.vip.vetbiz.gov/Default.aspx .
This notice is for planning purpose only and does not constitute a commitment on the part of the Government to award a contract, nor does the Government intend to pay for any information submitted as a result of this notice. This RFI is part of a continuous process for obtaining the latest information on the commercial status of the industry with respect to their current and near-term abilities. The information provided herein is subject to change and in no way binds the Government to solicit for, or award, a competitive contract. It is emphasized that this is a notice solely for planning and informational purposes
2.0 SCOPE
The scope of this effort falls in line with VA s mission, goals, and objectives in support of providing Veterans care. AED is a medical device that analyzes the heart s rhythm, if necessary; it delivers an electrical shock, known as defibrillation, which helps the heart re-establish an effective rhythm. It is a portable electronic device that automatically diagnoses the life-threatening cardiac arrhythmias of ventricular fibrillation and pulseless ventricular tachycardia in a patient, and is able to treat them through defibrillation, the application of electrical therapy which stops the arrhythmia, allowing the heart to reestablish an effective rhythm. An AED is the only effective treatment for restoring a regular heart rhythm during sudden cardiac arrest and is an easy to operate tool for someone with no medical background. The following Contract Line Items will be awarded in the proposed contract:
Contract Line Items
Manufacture
Description/
Part Number
0001
Zoll
AED Plus with AED Cover
20100001102011000
0002
Zoll
Fully automatic AED Plus with AED Cover
20100000102011000
0003
Zoll
AED PRO ZOLL Semi-Auto/Manual w/Lithium Bat & CPR-D Padz
90110320499991000
0004
Zoll
CPR-D padz (One piece Defibrillator Pads)
8900-0800-01
0005
Zoll
pedi padz II Defibrillator Pads
8900-0810
0006
Zoll
CPR stat padz HVP Multi-Function CPR Electrodes
8900-0402
0007
Zoll
ECG Electrodes - 3 ECG electrodes/pouch (200 pouches/600 electrodes)
8900-0003
0008
Zoll
SurePower Rechargeable Lithium Ion Battery Pack
8019-0535
0009
Zoll
SurePower Charging Station 4 charging bays
8050-0030
0010
Zoll
SurePower Single bay charger (ZOLL lithium batteries only) does not support lead acid
8200-000100
0011
Zoll
AED Pro non-rechargeable lithium batttery pack
8000-0860
0012
Zoll
AED Pro Soft Carry Case
8000-0810
0013
Zoll
CPR stat padz HVP Multi-Function CPR Electrodes (Case of 8 Pair)
8900-0400
0014
Zoll
Recessed Wall Mounting Box
8000-0814
0015
Zoll
Surface Mount Wall Cabinet
8000-0817
0016
Zoll
ZOLL Replacement Softcase
8000-0802
0017
Zoll
AED Pro Molded Vinyl Carry Case with Spare Battery Compartment
8000-0832
0018
Zoll
Standard Metal Wall Cabinet - Alarmed
8000-0855
0019
Zoll
AED Pro Hard Case with Foam Cut-Outs
8000-0875
0020
Zoll
Simulator/Tester
8000-0819
0021
Zoll
Real CPR Help Travel Trainer
8008-0006
0022
Zoll
AED Pro Simulator
8000-0829
0023
Zoll
CPR-D Demo Pad
8900-5007
0024
Zoll
AED Plus Demo Manikin.
8000-0835
0025
Zoll
Training CPR stat padz
8900-0190
0026
Zoll
AED Demo Kit.
8000-0834
0027
Zoll
AED Plus Trainer 2 Unit
8008-0050-01
0028
Zoll
CPR-D padz Training Electrodes
8900-0804-01
0029
Zoll
ZOLL Training Electrodes stat padz II, AEDPlusî, 6 pairs/case
8900-0805-01
0030
Zoll
Replacement Adhesive gels for CPR-D padz
8900-0803-01
0031
Zoll
Zoll CPR-D padz (One piece Defibrillator Pads)
8900-0800
0032
Zoll
Type 123 Lithium Batteries, quantity of ten (10) with storage sleeve
8000-0807-01
The VA is seeking vendors who can provide ZOLLî brand of AED as listed above or equal commodities which meet all of the following salient characteristics:
SC #
APPLICABLE CLINS
SALIENT CHARACTERISIT
METHOD OF EVALUATION
1
0001 - 0003
Must measure chest compression with audio and visual feedback that guides rescuers to deliver high-quality CPR.
Literature Review
2
0001 - 0003
Must simplify pad placement steps and guiding through verbal cues or diagram
Literature Review
3
0001 - 0003
Automatic energy selection (50J. 100J, 120J, 150J, 200J, and 250J)
Literature Review
4
0001 - 0003
Protected ECG circuitry
Literature Review
5
0001 - 0003
Patient connections are electrically isolated
Literature Review
6
0001 - 0003
Charge Time Less than 5 seconds 10 seconds
Literature Review
7
0001 - 0003
Defibrillation evaluates Voice Prompts guide: electrode connection and patient ECG determine if defibrillation needed
Literature Review
8
0001 - 0003
Data Recording and Storage ECG and CPR data
Literature Review
9
0001 - 0003
Audio recording installed with Minimum 15 minutes of audio recording
Literature Review
10
0001 - 0003
CPR: Metronome Rate: Variable 60 to 100 CPM
Literature Review
11
0001 - 0003
PC Minimum configuration and patient data recovery: Windows; 64MB RAM. VGA monitor or better. CD-ROM drive. IrDA port. 2MB disk space
Literature Review
12
0001 - 0032
Environmental Operating Temperature: 32ð to 122ðF; 0ð to 50ðC
Literature Review
13
0001 - 0032
Storage Temperature: -22ð to 140ðF; -30ð to 60ðC
Literature Review
14
0001 - 0032
Humidity: 5% to 95% relative humidity, non-Condensing
Literature Review
15
0001 - 0003
Vibration: MIL Std. 810F, Min, Helicopter Test
Literature Review
16
0001 - 0003
Shock: IEC 68-2-27; 100G: Described in MIL-STD-810E, Method 516.4 (shock)
Literature Review
17
0001 - 0003
-300 to 15,000 ft.; -91 m to 4573 m
Literature Review
18
0001 - 0003
Particle and Water Ingress: IP-55: IEC 529 Level 3 (IP3X) (spraying water) enclosure protection.
Literature Review
19
0001 - 0003
Weight - Complete unit, excluding batteries, shall not exceed 6.7 pounds
Literature Review
20
0001 - 0003
Dimensions Aid storage and portability, overall dimensions of unit shall not exceed 5.25 x 9.50 x 11.50 (13 cm x 24.1 cm x 29.2 cm)
Literature Review
21
0001 - 0003
Construction - Constructed to withstand the standard operating in a physician s office and/or Hospital, also withstand harsh operating conditions ambulance use and air transportation.
Literature Review
22
0001 - 0003
Carrying case with handle shoulder strap must be available and accessory storage compartments.
Literature Review
23
0001 - 0003
Service Life shall have a service life of not less than five years (excluding routinely replaced items such as a battery, accessories, etc.)
Literature Review
24
0001 - 0003
Warranty - Parts and labor shall be furnished under warranty for five full years.
Literature Review
25
0001 - 0003
Aligned with Emergency Cardiac Care (ECC) Guidelines
Literature Review
26
0001 - 0003
Meets OSHA Guidelines
Literature Review
27
0001 - 0003
Approved by the Red Cross Scientific Advisory Council
Literature Review
28
0001 - 0032
Must be new and not refurbished or remanufactured
Literature Review
3.0 RFI SUBMISSION INSTRUCTIONS
Companies are encouraged to respond if they have the capability and capacity to provide Zollî brand name or equal AEDs which meet all of the salient characteristics. However, be advised that generic capability statements are not sufficient for effective analysis of respondents' capacity and capability to provide the required product.
If your firm has an interest in responding to the requirement described above, please provide the following:
Corporate Capability Statement Information:
Corporate Capability Statement; demonstrates capabilities, knowledge and experience related to the manufacturing of the AEDs described above (please detail if capability exists to meet each SC listed). If significant subcontracting or teaming is anticipated in order to deliver technical capability and scale, organizations should address the administrative and management structure of such arrangements.
a) Provide a summary of your company s relevant experience in providing solutions that provide quick (real time) response to what if scenarios when schedule, technical changes, and financial changes occur with the program.
b) Demonstrate ability to provide a significant increase or decrease of support (e.g., rapidly identify, equip and employ resources to support the Government in the event of an immediate surge or reduction in requirements.)
Responses shall be:
Identified with the RFI number; and include: Company s name, CAGE code, mailing address, e-mail address, telephone and fax numbers, website address (if available), and the name, telephone number, and e-mail address of a point of contact having the authority and knowledge to clarify responses with Government representatives. Also please include Small Business designation if applicable.
Interested parties shall submit a Corporate Capability Statement not exceeding 10 single-sided, type-written pages using 12-point type or larger.ÃÂ
The page limit does not apply to advertising material such as company brochures and catalogs. If there is any information you deem proprietary, please mark accordingly.
SUBMISSIONS ARE DUE 26 OCT 2017, by 2:00PM, Eastern Standard Time. All responses under this RFI must be emailed to
[email protected].
Acknowledgement of receipt of responses will not be made, nor will respondents be notified of the OAO/SAC view of the information received. Do not send any material that requires a non-disclosure agreement or identify information that is business sensitive. Responses to this notice will not be returned. Telephone calls/questions or inquiries will not be accepted