36C25221Q1024 for Clinical Training Software as a Brand Name or Equal requirement,
Ventilation Mode Requirements:
Must be capable of the following ventilator modes:
–Volume-controlled Ventilation (VCV)
–SIMV-VC
–Pressure-Controlled Ventilation (PCV)
–Pressure Controlled, Volume Guaranteed (PCV-VG)
–SIMV-PC
–SIMV-PCVG
–Bi level Airway Pressure Ventilation (APVR capable)
–PSV-VG
–Non-Invasive Ventilation (NIV) and NIV-ST
–CPAP/Spontaneous
-Closed loop mode of ventilation to provide automated lung protective ventilation strategies to spontaneous and passive breathing patients.
Patient Monitoring Alarms:
Patient Monitoring Alarms:
-Low/high minute volume
- Low/high pressure
- Low/high tidal volume
- Low/high respiratory rate
Apnea time
- Low/high PetCO2
- Low/high pulse
- Low/high SpO2
- Low/high SpMet
- Low/high SpOC
- Percent leak
-PI (perfusion index)
-Humidifier water level
-Humidifier temperature
- Humidifier tilt
-Disconnection
-Oxygen alarm limit exceeded
-Oxygen concentration
-Disconnection
- Loss of PEEP
Exhalation obstruction
- HLI
-High PEEP alarm
- Flow sensor alarms
-ASV/APV
- CO2
-Power supply
- Gas supplies
-Cuff leakage
Ventilator Display Requirements:
-Real-time visualization of the lungs with representations of tidal volume, lung compliance, resistance and patient activity, including hemodynamic status and cuff pressure
-Visual representation of ventilator dependency and weaning process, grouped into oxygenation, CO2 elimination, patient activity, PetCO2 , SpO2 , pulse, and heart lung index (HLI)
-Graphic display of target and actual parameters for tidal volume, frequency, pressure, patient activity and minute ventilation
Display of more than 50 monitoring parameters
-Up to 8 waveforms at a glance based on: Paw, Pes, Paux, Volume, Flow, FetCO2 , PetCO2 , SpO2
-Up to 4 loops at a glance based on: Volume, Flow, Paw, Pes, Paux, Ptranspulm, SpO2 , CO2 , reference loops
-Simultaneous display of up to 17 parameter trends, selected from all monitoring parameters, for 1, 3, 12, 24, or 96 h
-Storage and display of up to 1,000 events with date and time stamp Graphic curve freeze and cursor function, inspiratory and expiratory hold; user configurable default graphics layout
Ventilator Features/Procedure Requirements:
-Suction Maneuver
-Automatic patient detection
-Screen snapshot
- Spirometry
- Integrated Medication nebulization including and ability to set phase cycle and duration of nebulization
- Integrated Aerogen nebulizer
-Automatic pulmonary mechanics measurements that include static compliance, inspiratory/expiratory resistance, auto PEEP and inspiratory/expiratory time constants
-Integrated capnography with volumetric CO2 measurement
-Integrated pulse oximetry
-Automated ET and Trach tube cuff pressure maintenance and monitoring
- Automatic maneuver for static compliance assessment and lung recruitment including transpulmonary pressure.
- Esophageal pressure monitoring and automated calculation of end inspiratory transpulmonary pressure/ end expiratory transpulmonary pressure
-Uses a proximal flow sensor to precisely measure the pressure, volume and flow directly at the patient’s airway opening.
-High flow oxygen therapy
-Heliox therapy
Ventilator Communication Requirements:
-USB and Compact Flash for screenshots
Updates and log data
- DVI with VGA output
-2 serial RS-232 ports (data monitoring / humidifier control)
- Special data interface for nurse call and alarm output: SpO2 & CO2 sensor, Aerogen and Cuff interface
Patient Types requirement:
-Adult
-Pediatric
-Neonatal