Intent to Award Brand Name Evo Tech Endoscope Cleaner and Reprocessor
10th Mission Support Group, Directorate of Contracting, United States Air Force Academy Colorado, intends to award a brand name only commercial purchase order in accordance with FAR Parts 12 & 13. This notice document and incorporated provisions and clauses are those in effect through Federal Acquisition Circular 2005-14, 22 Nov 06. The North America Industry Classification System Code [NAICS] is 423450 and the size standard is 500 employees. The requirement is to purchase: 1 each Product number 100012-01 Evo Tech⢠GI Suite (1 each 50004,20398, 22591, 20302, 13004001), 1 each Part Number 09003104 CH SEP, OLY>=40 series, 1 each Part Number 09601103 SCOPE CON SET, OLY GI>=40. These items will require identification adhering to DFARS 252.211-7003 (http://farsite.hill.af.mil). The justification for the brand name requirement is as follows:
a. This ECR is the only system with FDA clearance to automate manual cleaning. The Evo Tech Systemâs 33 minute cycle time includes the normally labor-intensive manual cleaning of the scope, which takes 6-10 minutes per scope. Additionally, this system automatically detects leaks thereby eliminatiing the inconsistencies associated with manual inspection.
b. The Evo Tech ECR System has a large basin size (21 inches) this helps to prevent endoscope coiling, there potentially reducing endoscope repair costs.
c. This system has an integrated MEC monitoring which assures precise results by eliminating the use of test strips. The single use of CIDEX OPA Concentrate Solution safely delivers fresh and consistent single metered dose of high level disinfectant every cycle.
Technical Data of the EvoTech Endoscope Cleaner and Reprocessor System:
Where marketed United States, Canada, New Zealand, Australia, and Singapore
FDA Clearance Yes
Configuration Floor, Stand-alone
Mobility Moveable, 4 casters, 4 levelers
Standard of Care High Level disinfection
Air Capability Yes, to all internal channels
Self-Disinfect Cycle Yes
Overall Size
Height (with lid closed) 117 cm (46 in)
Height (with lid open) 150 cm (59 in)
Width (without printer) 95 cm (37 in)
Depth 82 cm (32 in)
Weight (when empty) 226.8 kg (500 lbs)
Operating Weight Approximately 254 kg (560 lbs)
Basin Size 46.4 cm x 50.2 cm x 14.6 cm (18.25 in x 19.75 in x 5.75 in)
Service Access Area 36 total inches needed on one side. Unit can be moved for service.
Earthquake Restraint Unit includes attachment points
Basins per System 2
Scopes per Basin 1
Scope Compatibility Submersible flexible fiber optic and video endoscopes suitable for high-level disinfection
Cycle Time 33 minutes
Germicidal Agents
Approved Agents CIDEX® OPA Concentrate Solution
Water Supply
Hardness Water less than 50 PPM (CaCO3 Equivalent)
Type Reverse Osmosis (RI) and Deionized (DI) water cannot be used in the EvoTechTM ECR System.
Filter 0.2 micron bacterial filtered water
Water Pressure 40 PSI (275 kPa) minimum dynamic pressure during flow rate of 6 liters per minute per machine
Maximum static pressure is 100 PSI for the EvoTech System. Check water filtration manual for
maximum water filtration static pressure.
Water Temperature 5º C to 30º C (41º F to 86º F)
Please note: Hot water is not required. A water heater is included in EvoTech ECR System
Thermal Expansion Tank Included, wall mounted. Please note: Facility must install brackets for the thermal expansion tank prior
to installation of the EvoTechTM ECR System.
Detergent
Type CIDEZYMETM GI Enzymatic Detergent
Injection Yes, automatic
Alcohol
Reservoir Size 0.5 liter
Percentage 70%
EvoTechTM Endoscope Cleaner and Reprocessor (ECR)
Information Input/Output
Display 6.4-inch Color VGA
Hard Copy Printout
Control Interface Microprocessor
Input Touch Screen
Power Specifications
AC Power 208 V, 3~, 60Hz, 30 Amp. USA, Canada
220-230 V, 50Hz, 50 Amp. Singapore
240 V, 50Hz, 50 Amp. Australia
220-230 V, 50Hz, 50 Amp. New Zealand
Power outlet must be ground fault protected. A recommended ground fault circuit interrupter (GFCI) for
3-phase 208V system includes 3-phase Delta connection with no neutral and Auto Power Recovery on
power loss.
Electrical Connection 3-phase NEMA, L 15-30
Printer Connection 90-240 V
Power Cord Length Approximately 6 feet
Drain Specifications
Number of Drains The left and right basins each has a drain tube. There is a shared overflow detector
Drain Capacity 25 liters/min. per drain
Maximum Temperature 75° C (167° F)
Drain A 38 inch high (minimum) 42 inch high (maximum), from the floor, vented stand pipe, 2.5 to 3 inches in
diameter, or 2 each 1.5 inch diameter, vented stand pipes must be in place for the waste discharge
hoses. The pipes MUST NOT be capped. One end of the discharge hose should be connected
(permanent connection) to the drain hose barb fitting on the EvoTechTM ECR System. The other end
is inserted into the stand pipe. Please consult your local water municipality for any region-specific
requirements.
Operating Conditions
Min. Room Temperature 15° C (59° F)
Max. Room Temperature 30° C (86° F)
Air Exchange Rate 12 times per hour
Relative Humidity 10 - 80% non condensing
Elevation -100 to 3,000 meters (-330 to 9,900 feet)
Tilt up to 1 degree from vertical
Heat Load
Normal Cycle 1200 Watt (4000 BTU/hr)
Self Disinfection Cycle 2000 Watt (7000 BTU/hr)
Idle 300 Watt (1000 BTU/hr)
Printer Information
Printer Black and white printer included. Should be GFI protected.
Printer Support The printer rests on either side of the system on the provided support (brackets are included). The
printer can also be located on a nearby shelf. Printer brackets can fit up to a 195 mm x 195 mm printer.
Printer brackets extend approximately 220 mm from the system.
Printer Connection Serial cable
CE Mark Yes
Configuration Wall-mounted
Self-Sanitize Cycle Yes
Filter Media
1.0 micron Filter Polypropylene depth media
0.2 micron Filter Polypropylene polysulfone
Water Supply
Incoming Pressure 2.76 to 4.14 bar (40 to 60 PSI) Maximum pressure shall not exceed 5.52 bar (80 PSI)
Pressure Differential
1.0 micron Filter 3.33 bar (50 PSI)
0.2 micron Filter 2 bar (30 PSI)
Temperature 5º C to 30º C (41º F to 86º F)
Power Specifications
Voltage 100, 120, 230, 240 VAC (selectable), 50/60 Hz single phase
Water Filtration System should be GFI protected.
Fuses 100-120 VAC, 1 4A 3AG (1.25 in x 0.25 in)
220-240 VAC, two T2A (5 mm x 20 mm)
Light Bulbs 12 V incandescent with slide base
Overall Size 706.1 mm W x 701 mm H x 165.1 mm D (27.8 in W x 27.6 in H x 6.5 in D)
Weight Wall mounting must support 44.5 kg (100 lbs)
Filled with water ~16 kg (35 lbs)
Water Filtration Plate Mounting brackets are required to be installed by facility prior to ASP Water Filtration System
(Mounting Plate) installation
Description The Water Filtration System from Advanced Sterilization Products is designed to filter out
sediment found in municipal water sources and to provide bacteria-free and reduced
pyrogen water for use in endoscope reprocessing. The convenient wall-mounted design
includes a 1.0 micron filter to remove larger particles and sediment, and a 0.2 micron filter
to remove fine particles and bacteria from the incoming water source.
All responsible sources may submit a response, which if timely received, will be considered by 10MSG/LGCB. The Government does not intend to pay for any information provided under this presolicitation notice.
Point of Contact is Sam Richmond, Contract Specialist (719) 333-3948 email: (
[email protected]) or alternate Leslie Ryan, Contracting Officer (719) 333-3961 email:
[email protected].
5352.201-9101 OMBUDSMAN (AUG 2005) (a) An ombudsman has been appointed to hear and facilitate the resolution of concerns from offerors, potential offerors, and others for this acquisition. When requested, the ombudsman will maintain strict confidentiality as to the source of the concern. The existence of the ombudsman does not affect the authority of the program manager, contracting officer, or source selection official. Further, the ombudsman does not participate in the evaluation of proposals, the source selection process, or the adjudication of protests or formal contract disputes. The ombudsman may refer the party to another official who can resolve the concern. (b) Before consulting with an ombudsman, interested parties must first address their concerns, issues, disagreements, and/or recommendations to the contracting officer for resolution. Consulting an ombudsman does not alter or postpone the timelines for any other processes (e.g., agency level bid protests, GAO bid protests, requests for debriefings, employee-employer actions, contests of OMB Circular A-76 competition performance decisions). (c) If resolution cannot be made by the contracting officer, concerned parties may contact the USAF Academy Ombudsmen, Lori Staples 8110 Industrial Drive, Ste 103 USAFA, CO 80840 Telephone number 719-333-3907 FAX 719-333-9018 email:
[email protected] Concerns, issues, disagreements, and recommendations that cannot be resolved at the MAJCOM/DRU level, may be brought by the concerned party for further consideration to the Air Force ombudsman, Associate Deputy Assistant Secretary (ADAS) (Contracting), SAF/AQC, 1060 Air Force Pentagon, Washington DC 20330-1060, phone number (703) 588-7004, facsimile number (703) 588-1067. (d) The ombudsman has no authority to render a decision that binds the agency. (e) Do not contact the ombudsman to request copies of the solicitation, verify offer due date, or clarify technical requirements. Such inquiries shall be directed to the Contracting Officer. (End of clause)
Point of Contact(s):
Sam Richmond, Contract Specialist, Phone 719-333-3948, Fax 719-333-9103, Email
[email protected]
Bid Protests Not Available