Large
Specifications:
16 sq. ft. interior growing area
56 inches of growing height
RH range: 40-95%
Transparent barrier between the lamps and growth area for more precise temperature control in the growing area.
CO2 control, scrub and supplementation, to provide both below-ambient and above-ambient CO2
extended cold temperature down to -2 C.
Light: Max > 600 micromole/m^2/S, adjustable light levels Flexibility of using various lights for desired light quality
Air exchange rate: < 1/hr (less than one air exchange per hour) Air sample port: for continuous air quality monitoring
Sensor output interface: allows data transmission between inside and outside of the chamber. Sensors are located inside of the chamber. Data can be collected and monitored without opening the chamber. Data is logged on an external computer.
Irrigation line: allows watering plants inside using external water supply
Integrated computer control of multiple growth chambers.
Day/night temperature, humidity, CO2, light set points
Small
Specifications:
29 cu. ft. interior growing area
Min. 50 inches interior height
2 shelves, adjustable, with separate light canopy for each
Min. 400 umol/m2/s PAR light intensity per shelf
Adjustable light intensity for each shelf
RH range: 40-90%
Temperature: 4-40 C
Stepwise and diurnal temperature control options
CO2 control: scrub and supplementation (100 - 2000 umol/mol)
Data monitoring and logging of temperature, light, and CO2
Observation window
Castors for mobility