This is a request for information only and in no way, shape, or form, binds
the government to any contract or agreement
Requirement: Brand name FT-IR Spectrometer, conforming to the specifications of the Nicolet iS50R FT-IR Spectrometer.
- Must have a Vectra-Plus Interferometer.
- Must have all optics aluminum coated.
- Must have a measurement response from the Ultraviolent (UV) through Far Infrared (FWIR) spectral band.
- Must have better than or equal to 0.09 cm-1 Spectral Resolution.
- Must have better than or equal to 0.07%T Ordinate Linearity.
- Must have better than or equal to 0.01cm-1 Wavenumber Precision.
- Must have a Spectral Measurement range better than or equal to 350cm-1 to 27,000cm-1.
- Must have a Scan Velocity better than or equal to 0.158-6.28cm/sec.
- Must have Rapid Scan performance better than or equal to 65 or 95 spectra per second.
- Must have Polaris High Stability IR Source with a Spectral Region better than or equal to 9600cm-1 through 20cm-1.
- Must have Tungsten-Halogen VIS/NIR Source with a Spectral Region better than or equal to 27,000cm-1 through 2000cm-1.
- Must have internal storage with automatic selection of Beam Splitters.
- Must have an internal automatic filter wheel and polarizer.
- Software Development Kit (SDK) must be available for a software application that communicates and controls measurement function required for external measurements.
- Must have better than or equal to 55,000:1 SNR for a 1 minute scan.
- Must have better than or equal to 13,000:1 SNR for a 5 second scan.
- Must be configurable for both internal and external source beam path.
- Must be configurable for both internal and external emission beam path.
- Must be configurable to use either internal or external measurement detectors.
- Must have the appropriate detectors to cover the broad range of spectral measurements.
- Must be a sealed and desiccated unit providing protection from dust and moisture.
- Must have the ability to purge the internal optical path.
- The Spectrometer dimensions must be close to 25”x27”x11” (W x D x H).
- The Spectrometer weight must be close to 132 lbs.
- An application PC must be provided and be compatible with the Windows10 operating system.
- Include performance and checkout data with system when delivered.
- System and component delivery should be within 90 days ARO.
- Package and mark system and components in an industry standard way for shipment of such fragile instruments.
- Select shipper that can and will appropriately handle delicate instrument shipments.
- Provide on-site service technician as necessary for system unpacking, installation and checkout.