Kansas City VA Medical Center
4801 E. Linwood Blvd.
Kansas City, MO 64128
Audiology
SCOPE OF WORK AND GENERAL SPECIFICATIONS FOR:
HEARING INSTRUSTMENT VERIFICATION SYSTEM
1. Background
The Lenexa CBOC is opening in June 2021 and Audiology is asking for two hearing instrument verification systems to be placed in each hearing aid fitting room.
2. Specifications
DESCRIPTION: HEARING INSTRUMENT FITTING SYSTEM
MUST MEET OR EXCEED SALIENT CHARACTERISTICS BELOW:
REQUIRED FEATURES:
Dual probes for fast binaural fitting
Viewport featuring Speechmap, patented directional test, unique noise reduction and feedback suppression tests.
Test signal for frequency manipulating instruments
Fitting protocols (DSL 5.0, NAL-NL1, Camfit)
True two channel measurements
Single/Dual view display options
External monitor capability
Sensory Loss Simulator
ANSI, ITE & BTE couplers
Occlusion effect test
Battery Stimulator
RECD transducer
Probe monitor
Telecoil test
Manual Gain, output, and distortion tests
DIMENSION REQUIREMENTS:
Overall dimensions
Rear main unit: 14.5 x 16 x 6.5in
Test Chamber: 14 x 14.4 x 4in
Weight: 17.6 lbs
UTILITY REQUIREMENTS:
Power consumption: 100-240V, 50-60Hz, 250VA
Temperature: -20 to +60C
Relative humidity (non-condensing): 5 95%
GENERAL REQUIREMENTS:
Display type: fluorescent backlit active color
Display size: 12.1 diagonal
Resolution: 800 x 600 (SVGA)
Internal Printer: 3 thermal line printer, 200 dots/inch
Headphone amplifier : 2 @ 5 watts each
Stimulus channels: 2
Measurement channels: 2
Connectors:
1 ethernet (RJ45)
4 USB
1 HDMI
2 external speakers
2 expansion jacks (1/4 mono)
2 real ear mics (3.5mm st)
1 RECD transducer (3.5mm st)
1 monitor headset (3.5mm st)
1 test chamber ref. mic (3.5mm st)
1 coupler mic (3.5mm st)
1 battery substitute (3.5mm st)
TEST BOX REQUIREMENTS
Working space: 11 x 5 x 1.5in
2 independent speakers 2 x 3in
Test Box isolation @1K Hz: .25dB
Induction Coils per ANSI S3.22-2003: 1 telephone magnet field sim (TMFS) & 1 23cm diameter test loop
Battery Simulator: per ANSI 3.22-2003
Frequency Range: 200-8000Hz
Test Stimuli: Tone, Tone Burst, Pink Noise, Dual Direction Noise, Calibrated or Live Speech, ISTS
Test Stimulus Levels: 40 to 90dB in 5dB steps
Test Stimulus Levels (inductive): 31.6mA/m per ANSI S3.22 - 2003
Test Stimulus Distortion: <2% at 90B SPL and 0.5% at 70dB SPL
Test Stimulus Accuracy at Reference Mic for tones (200-2000Hz)+1.5dB SPL
Test Stimulus Accuracy at Reference Mic for tones (2000-8000Hz)+2.5dB SPL
Equalization Method: Real time modified pressure method (stored for open fittings)
Analysis frequencies per octave: 12
Analysis filter bandwidth (noise): 1/12 octave
Measurement accuracy at 1kHz for tones: +/-1dB
Measurement accuracy re 1kHz for tones: +/-1dB (200-5000Hz) and +/-2.5dB (5000-8000Hz)
Measurement Range: 30-140dB SPL
Harmonic Distortion measurement: 2nd and 3rd or 2nd plus 3rd
Harmonic distortion range: 200-4000Hz
Harmonic distortion accuracy: +/-1%
Battery Drain range: 0-20 mA
Battery Drain accuracy: +/-5%
Battery Drain resolution: +/-.01 mA
ANSI S3.22 1996 and 2003 tests available: OSPL90 Full-on Gain Reference Test Gain Frequency Response Frequency Range Maximum OSPL90 Harmonic Distortion Attack & Release Time Equiv. Input Noise Input/Output Curves Coupler SPL-Telephone Simulator Simulated Telecoil Sensitivity Battery Drain
OTHER Tests Available: Speechmap* real ear audibility measures Real-Time adaptive directional verification Coupler SPL vs. frequency Coupler gain vs. frequency Spectral Analysis Noise reduction verification Distortion vs frequency Manual measurement of output, gain, and distortion VIEWPORT audibility, directional, noise reduction, and feedback suppression verification.
ON-EAR MEASURES
Speakers: 2 2x3 in. (5x7.5cm) ducted ports
Probe microphone tube: Silicone 1mm diameter x 75mm
Probe modules: 2 each containing probe and reference microphones
Probe microphone noise floor (200-8000Hz): <45dB SPL
Frequency range: 200-8000Hz
Test Stimuli: frequency modulated tone, tone burst, pink noise, dual directional noise, calibrated speech, live speech, ISTS
Frequency modulation: sawtooth +/-3% over 128ms
Test stimulus levels at reference microphone for tones at 1m: 40-90dB SPL in 5 dB steps
Test stimulus accuracy at reference microphone for tones (200-2000Hz): +/-1.5dB SPL
Test stimulus accuracy at reference microphone for tones (2000-8000Hz): +/-2.5dB SPL
Equalization method: real time modified pressure method (stored for open fittings)
Analysis frequencies per octave: 12
Frequencies per octave (tone burst): 3
Analysis filter bandwidth (speech, noise): 1/3 octave
Measurement accuracy at 1kHz for tones: +/-1dB
Measurement accuracy at 1kHz for tones: +/-1dB (200-5000Hz)
Measurement accuracy at 1kHz for tones: +/-dB (5000-8000Hz)
Measurement range: 20-135dB SPL (200-2500Hz) & 30-140dB SPL (2500-8000Hz)
ANSI S3.46 1997 TESTS AVAILABLE
Real Ear unaided Response Real Ear Aided Response Real Ear Occluded Response Real Ear Insertion Gain
OTHER TESTS AVAILABLE
Speechmap* real ear audibility measures Real-Time adaptive directional verification Coupler SPL vs. frequency Coupler gain vs. frequency Spectral Analysis Noise reduction verification Distortion vs frequency Manual measurement of output, gain, and distortion VIEWPORT audibility, directional, noise reduction, and feedback suppression verification.
FITTING METHODS AVAILABLE
Speechmap with DSL 5.0a, NAL-NL1, CAMFIT, Insertion gain with NAL-RP, NAL-NL1, Fig 6, PogoII, Berger, Libby